The Speech Processor Test Device
is a troubleshooting tool for detecting the most common
audio processor problems. To troubleshoot
with the Speech Processor Test Device, switch on your SONNET. Place the flat side of the DL-Coil
under the Speech Processor Test Device. Speak into the SONNET microphones. The indicator light should flicker
to the rhythm of your voice. If it does not, adjust the volume
and speak into the microphones. If the indicator light still does not flicker, change the batteries
and speak into the microphones. If the indicator light still does not flicker,
replace the coil cable. If your SONNET still is not working
or for immediate assistance, please contact your CI centre.